25
At any stage of the treatment setting mode, holding the FUNC button for
approximately 5 seconds allows switching to Set-Up Menu.
5.1.1. Risk of Bleeding Evaluation
When entering the treatment setting menu, the following screen is displayed:
Insert printout
The default risk of bleeding value is flickering.
This function determines the evaluated risk of bleeding from the wound.
The following risk evaluations are available: High, Moderate or Low risk of
bleeding. As a default, for the safety of the patient, the risk is pre-set to be
High. For each option, the working neg. pressure will be limited as follows:
Risk of Bleeding
Working
Neg. Pressure [mmHg]
High
Moderate
Low
Intermittent Operation Mode
30-65
30-90
30-200
Cyclic-Cont. Operation Mode
50
40 / 50 / 75
40 / 50 / 75 / 100 /
125
As the Risk of Bleeding is set to High or Moderate, for the safety of the
treatment, Max Flow and Alarms will be limited to the default values of 100
ml/hr and Audio & Visual, respectively.
Use the up and down scroll arrow buttons in the control panel to change the
flickering value and determine the evaluated Risk of Bleeding.
Press the OP/SET button to finalize your selection. The value stops flickering.
Press ↓ to continue to the next parameter.
Warning: Consider carefully any change from the default settings.
5.1.2. Use Previous Settings
The user may choose to use the previous treatment settings if the Risk of
Bleeding is identical to that of the previous treatment. The default value of this
function is NO.
Use the up and down scroll arrow buttons in the control panel to change the
displayed value and determine whether or not to use the previous settings.