42
Auditory alarm signal: low frequency repetitive beeping sound.
Additional signals: general warning status (symbol no. 7) from screen symbols
description table.
Information sentences: First line:
“Ineffective vacuum level”, Second line: (1) “Check
for leakage” (2) “and tubing connectivity” (where (1) and (2) are presented one after
the other for 5 seconds each).
To resolve this alarm:
•
Inspect the sponge, drape strips, tubing and collection canister for possible
leakage.
•
Make sure that the canister is positioned properly in the canister socket.
•
If connected to an external vacuum source, make sure that the pressure from
the external vacuum source is above the working pressure.
•
In case the applied vacuum is ~0 mmHg but the pump is working, contact your
service provider for hydro
-gel filter inspection.
In case this alarm is repeated after following these instructions, contact your service
provider.
Warning: Do not leave a wound covered without effective vacuum for long time
periods.
Warning: failure to respond to this alarm may exacerbate wound infection.
1.7.
Low Battery alarm
Description: this alarm is initiated when battery voltage is low and approximately 10
minutes of operation are left.
Priority: Low.
Visual alarm signal: constant yellow indication LED, alarm signal no. 1- general
alarm signal from screen symbols description table.
Auditory alarm signal: low frequency repetitive beeping sound.
Additional signals: general warning status (symbol no. 7), low battery status (symbol
no. 8 with 1 black bar) from screen symbols description table.
Information sentences: First line:
“Low battery”, Second line: “connect to main
source”.
Summary of Contents for Vcare a Series