Length: 1.7 meters
The external vacuum source connector can tolerate vacuum of up to 760 mmHg;
however, the recommended input vacuum is max. 200 mmHg.
Internal fuse
Quick blow fuse 5A / 250V
Buzzer
Manufacturer: Mallory Sonalerts products INC.
Sound pressure range @ 10 cm: 88 dB, @ 1 m: 70 dB.
High priority alarms
Specifications:
Beep
, Pause 100 ms,
Bee
p, Pause 100 ms,
Beep
, Pause 300 ms,
Beep
,
Pause 100 ms,
Beep
, Pause 500 ms,
Beep
, Pause 100 ms,
Beep
, Pause 100 ms,
Beep
,
Pause 300 ms,
Beep
, Pause 100 ms,
Beep
.
Repeat
every 2.5 sec’s.
Beep “on” time is 100 ms.
Low priority alarms
Specifications:
Beep
, Pause 200 ms,
Beep
.
Repeat every 5 minutes.
Beep “on” time is 200 ms.
Designed to Meet IEC 60601-1-8
External power adapter
The Vcare
®
unit can be connected to the main power supply through a power adapter.
The power adapter is not part of the Vcare
®
device but it is mandatory to use it. The
power adapter must comply with the specifications described below.
Input
Voltage: 100-240 V
Max. Power Consumption:
approximately
50 W
Max. input current: 1.5A
UL / TUV / CSA medical grade certified.
Output Voltage:
approximately
15V DC
Operation Temperature: ~0-65
C
Max. output current: 3.3 A
Operation relative humidity: ~20%-90%
Leakage current: 0.5mA / 240VAC
Storage Temperature: ~(-20)-85
C