29
The default value of the lower limit is pre-set to be 10 mmHg
below the working neg. pressure and the range of the lower limit
can be linearly adjusted within the range of 30 mmHg to the
default value.
For advanced Lower Limit settings, press FUNC shortly. This
allows the user to change the Lower Limit to a different value in
the pre-set available range. Use the up and down scroll arrow
buttons in the control panel to change the displayed value and
determine the required lower limit.
Press OP/SET button to finalize your selection. The displayed
value stops flickering.
Press ↓ to continue to the next parameter.
Warning: do not set low values of lower limit in order for the
system to detect improper vacuum applied to the wound.
In case the pressure applied by the device will be lower than the
lower limit value, an alert will be generated (see the Alarms
chapter for more details).
5.1.3.2.3. Upper Limit
When entering the Upper Limit function, the following screen is
displayed:
Insert printout
The default upper limit value is flickering.
The default value of the upper limit is pre-set to be 15 mmHg
above the working neg. pressure and the range of the allowed
upper limit is pre-set according to the Risk of Bleeding and
Working neg. Pressure (WP) selection as follows:
Risk of Bleeding
Upper Limit Range
High
WP+15 to 80
Moderate
WP+15 to 105
Low
WP+15 to 215