59
communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of the
transmitter
m
150 kHz to 80
MHz
80 MHz to 800
MHz
800 MHz to 2,5 GHz
0.01
0.17
0.17
0.23
0.1
0.37
0.37
0.07
1
1.16
1.16
2.33
10
3.69
3.69
7.38
100
11.66
11.66
23.33
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection form structures, objects and people.
Definitions of symbols used
Warning for possible hazard to
system, patient or operator
Manufacturer
Important Operational Information
SN
Serial Number
Follow Instructions For Use
REF
Model Name
Not for General Waste
Direct Current
Fuse
Do not reuse
Type BF Applied Part
Authorized
Representative in the
European Community.
On
Do not re-sterilize
Keep dry
Off
Temperature Limits Symbol
This side up
Manufactory date
Batch lot number
Summary of Contents for Vcare a Series
Page 1: ...1 User Manual ...