Visual alarm signal: flickering red indication LED, alarm signal no. 1- general alarm
signal from screen symbols description table.
Auditory alarm signal: high frequency repetitive beeping sound.
Additional signals: general warning status (symbol no. 7), low battery status (symbol
no. 8 with no black bars) from screen symbols description table.
Information sentences
: First line: “Critical Battery Condition”, Second line: "Connect
to main source”.
To resolve this alarm: connect the Vcare
®
unit to the wall outlet using a suitable
external power adapter (see specifications section) and press OP/SET to resume
treatment.
Battery malfunction is expected if the system is not connected to an external power
adapter immediately.
In case this alarm is repeated after following these instructions, contact your service
provider.
1.2.
High flow alarm
Description: this alarm is initiated when the system detects that the accumulated
liquids volume in the collection canister exceeds the pre-defined maximal volume per
hour.
In this case, vacuum operation is automatically shut down and alarms are set.
High flow stops the vacuum operation.
It will not be possible to re-activate the device by pressing the OP/SET button.
For re-operation, the device will have to be re-started.
Priority: High.
Visual alarm signal: flickering red indication LED, alarm signal no. 1- general alarm
signal from screen symbols description table.
Auditory alarm signal: high frequency repetitive beeping sound.
Additional signals: general warning status (symbol no. 7) from screen symbols
description table.
Information sentences
: First line: “Watch for possible bleeding”, Second line: “Turn
off for reactivation”.