Error alerts
3.2.
Err01: Extreme High Vacuum.
Description: this alert is initiated In case Extreme High Vacuum alarm is repeated
more than 3 times is 24 hours.
Alarm signals: none.
Additional signals: general warning status (symbol no. 7) from screen symbols
description table.
Information sentences: First line: "Error Alert Err01 Postpone Done", Second line:
“Refer to user manual for details".
To resolve this alert:
•
If connected to external suction, make sure that external vacuum is 50 mmHg
above the working neg. pressure (which were determined) and check if this error
alert is repeated.
•
Reduce the external vacuum level to 10 mmHg above the working neg.
pressure.
•
?ינוציח ץחל עובקל היצפוא שי םילוח תיבב םאה
•
If the alarm is repeated, contact IVT Medical Ltd. certified technical personnel for
inspection of the pressure regulation valves.
3.3.
Err02: Flow LEDs error
Description: the canister fill level bars display is based on the internal canister fill
LEDs detection. The device automatically tests each one of the 16 flow LEDs
(sensors) functionality every time it is turned on. In case the LEDs self-test detects a
failure in one (or more) of the LEDs in two sequential operations, an error alert will
be displayed.
Alarm signals: none.
Additional signals: none
Information sentences: First line: "Error Alert Err02 Postpone Done", Second line:
“Refer to user manual for details.
To resolve this alert:
•
Check for correlation between the canister fill level bars on the screen display
and the actual liquid volume inside the canister (each bar represents 100 ml
of liquid). In case the display and the actual liquid volume are correlated, the
Summary of Contents for Vcare a Series
Page 1: ...1 User Manual ...