Event-Timer (1800.5) ready OR
•
Changing of measuring value >= Analog Input Delta Transmission Value (6133h),
if 6133.1 > 0
If the Transmission-Type is = 254, the Event-Timer is ignored.
If the Inhibit-Time (1800.3) is >0 and not expired yet, no further PDO will be transmitted.
Default settings
:
CAN-Bitrate
500kBits/s
Node-ID
0x40
Transmission-Type (Obj. 1800.2)
255
Inhibit-Time (Obj. 1800.3)
0, that means disabled
Event-Timer (Obj. 1800.5)
0x03E8, that means 1 PDO /s
Analog Input Delta Transmission Value
(Obj. 6133.1)
0, that means Delta-Interrupt-function
disabled
Producer Heartbeat Time (Obj. 1017)
0, that means Heartbeat disabled. The
Bootup-frame will be sent once after power-up
or after reset.
All other objects and settings: see EDS and ba-gsv2CanOpen.pdf
General advices and Tips
Please attend the actual documents on:
http://www.me-systeme.de/software.html
The command number is transmitted as a single byte to the measuring amplifier.
The command parameters and the return parameters are binary bytes and not numerical
(ASCII-) values (e.g. the parameter „01“ means 0x01 and not 0x31).
A Windows dynamic-link-library (MEGSV.DLL) to access all the features of the measuring
amplifier is available, also a LabView example VI with dll-function wrapper-VIs.
For direct programming via RS232 it is sometimes easier to use the text format for
measuring values, provided that the scaling factor was calculated correctly, according to
the application requirements, since in text format the GSV-2 itself does the multiplication
of the raw values with the scaling factor.
Otherwise, in the binary format every measuring value needs to be multiplied with the
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf Germany
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en