Possible error codes 0xA0,0x54,0x71
Get Digits
Command number: 62
Number of parameters: 0
Bytes sent by the GSV: 1
Get Digits determines
the number of digits shown in the LC-Display.
Possible error codes 0xA0,0x54
Get Channel
Command number: 65
Number of parameters: 0
Bytes sent by the GSV: 1
Get Channel
reads the the active analog input channel. In the standard version there are
two inputs available, the bridge input (channel 0) and the channel 1, which has an input
voltage range from 0 to 10V.
Possible error codes 0xA0,0x91
Get Last Error
Command number: 66
Number of parameters: 0
Bytes sent by the GSV: 1
With get last error the the error status of the last command given can be determined.
Every command – with the exception of
Get last error
itself – overwrites the error register.
That is useful to determine the reason for the rejection of a write command. The command
Reset status
(No. 0) resets the error register to 0.
The following
error codes
are actually defined:
Default: (no command given or error code cleared):
0x00
No Error (OK), no additional parameter changes made:
0xA0
No Error (OK), but* additional parameter changes made:
0xA1
From here on: command given was rejected because:
Command number not existent:
0x40
Command No existent but not implemented in this Firmware:
0x41
Access denied (unspecified):
0x70
Access denied, because "Blocking" is active:
0x71
Access denied, because password not given or incorrect:
0x72
Access denied, because configuration jumper is not set:
0x73
Access denied, because number of allowed executions is too high:
0x74
Access denied, because Set/Write prohibited at this port:
0x75
wrong parameter, (unspecified):
0x50
wrong parameter, (partly) wrong bits:
0x53
wrong parameter, parameter absolutely too big:
0x54
wrong parameter, parameter absolutely too small:
0x55
wrong parameter, invalid parameter combination:
0x56
wrong parameter, parameter relatively** too big:
0x57
wrong parameter, parameter relatively** too small:
0x58