Description of the keys
Key
Function
MODE
Switch on and off or enter the logger-menu
MENU
(LEFT)
Enter into main menu or cancel an input.
Move to upper menu level.
If changing number: Move cursor left
UP
Browsing the menu entries or augment a number / digit value.
DOWN
Browsing the menu entry or reduce a number / digit value.
OK
(RIGHT)
Confirm an input or change into a sub menu.
If changing number: Move cursor right.
SHORT
Connecting the Ud and -Ud (Short-circuit the sensor signal)
ZERO
Triggering an automatic zero-adjustment
Number format
To set the number value and the date or the time move with a short press of OK the
cursor to the right and with MENU the cursor to the left. The number above the cursor is
blinking and can be increased or reduced with UP / DOWN buttons.
I
n order for the setting to take effect, the OK button must be kept pressing until the
whole number flashes.
Then release the OK button, it appears “OK to set”. Confirm this with pressing on OK. To
cancel the time setting press the MENU button long time.
Settings of the measuring value display (scaling factor)
scaling factor = Input sensitivity / Nominal excitation * Nominal load
Input sensitivity of the measuring amplifier
e.g. 3,5 mV/V
Measuring range of the sensor
20 kN
Nominal excitation of the sensor
1,9998 mV/V at 20 kN
This example results in an scaling factor of 35,004.
RS 232/422 Protocol of the GSV-2 Strain Gage Bridge Amplifier GSV-2
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf Germany
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en
15