Example (bipolar):
Shall NoiseCut-Threshold be = 0x810000; then all measuring values, that are between
0x810000 and 0x7F0000, set to 0x800000 exactly.
Value range: bipolar: 0x800000..0xFFFFFF; unipolar: 0x000000..0x7FFFFF
possible error codes: 0xA0, 0x71,0x75, 0x54, 0x55, 0x84, 0x41
Get NoiseCut Threshold
(only if NoiseCut existent)
Command number: 149
Number of parameters: 0
Bytes sent by the GSV: 3
With
GetNoiseCutThreshold
you can read the register described before.
Possible error codes 0xA0,0x91, 0x41
Set AutoZero Counter
(only if AutoZero existent)
Command number: 150
Number of parameters: 2
Bytes sent by the GSV: 0
With Set AutoZero Counter the counter for the automatic periodic tare-function can be set.
If this AutoZero function is enabled (see SetSpecialMode, Bit 14), after <AutoZero-
Counter> number of mesuring values a zero-point adjustment will be executed
automatically, if the threshold switch 1 „SW1“ is off (meaning the measuring value is below
this threshold). The mesuring values are only counted, if the threshold switch 1 „SW1“ is
off. By using this function, an eventual drift of the measuring value (e.g. because of
temperature drift of the sensor) can be compensated. So, the time-period of this
automatic zero-setting depends on the AutoZero counter and the data rate.
Example (bipolar):
Shall the upper threshold 1 be =0x810000, the data rate = 10/s and the AutoZero-Counter
=100. Then every 10 seconds a zero-point adjustment will be executed automatically, if all
measuring values within this period of time were below 0x810000.
Value range: 0x0001..0xFFFF
Possible error codes: 0xA0, 0x71,0x75, 0x55, 0x84, 0x41
Get AutoZero Counter
(only if AutoZero existent)
Command number: 151
Number of parameters: 0
Bytes sent by the GSV: 2
With
GetAutoZeroCounter
you can read the register described before.
Possible error codes 0xA0,0x91,
0x41
Set UserTextChar
(devices with LC-Display only)
Command number: 152
Number of parameters: 2
Bytes sent by the GSV: 0