write offset
Command number: 8
Number of parameters: 2
Bytes sent by the GSV: 0
Write offset
sets the Offset configuration of the preamplifier. The bytes sent should
contain a value which was recently determined with
read offset
.
Remarks:
Only the commands
write offset
and
set offset
affect the analog output.
Affected register:
Offset.
Value Range: 0x00.00..0x0F.FF
possible error codes
:
0xA0,0x54,0x71
Get all
Command number: 9
Number of parameters: 1
Bytes sent by the GSV: 0
Get all
restores a set of configuration, (some might be stored previously with
Save All
),
chosen by the parameter:
0:
:
Restore Setting before the last power on cycle.
1:
:
Restore Manufacturer Settings
2...7
:
Restore configurations saved at dataset 1...6
Affected registers:
channel, gain, frequency, offset, zero, scale, threshold.
Value Range: 0x00..0x07
possible error codes: 0xA1,0x54,0x80,0x71
Save all
Command number: 10
Number of parameters: 1
Bytes sent by the GSV: 0
Save all
saves all relevant registers of the GSV to a configuration set in an internal
memory. This data remains even when the device is turned off. It is possible to save
different configuration sets. After turning on the amplifier up to 64 memory procedures are
possible. The parameter indicates the number of the configuration set in the memory.
Parameters= 2 to 7:
: configuration set 1 to 6, saved by the user.
The positions 0 and 1 can not be programmed by the user. On the position 0 the current
configuration of the GSV is automatically saved.
You can restore the configuration set with
get all
.
Value Range: 0x02..0x07
Possible error codes: 0xA0,0x54,0x55,0x74,0x71