![ME-Messysteme GSV-2AS User Manual Download Page 22](http://html1.mh-extra.com/html/me-messysteme/gsv-2as/gsv-2as_user-manual_1757596022.webp)
Read offset
Command number: 4
number of parameters: 0
Bytes sent by the GSV: 2
Read offset
determines the offset setting of the analog front end of the amplifier. The
value received can be send back with
write offset
.
Possible error codes: 0xA0, 0x91
Write scale
Command number: 5
number of parameters: 3
Bytes sent by the GSV: 0
Write scale
sets the input sensitivity of the AD converter. The 3 bytes sent should contain
a value which was recently determined with
read scale
.
Affected register:
Scale.
Value Range: 0x20.00.00..0xFF.FF.FF
possible error codes
:
0xA0, 0x55,0x71
Write zero
Command number 6
number of parameters: 3
Bytes sent by the GSV: 0
Write zero
sets the zero adjustment of the AD converter. The 3 bytes sent should contain
a value which was recently determined with
read zero
.
Affected register:
Zero.
Value range: 0x00.00.00..0xFF.FF.FF
possible error codes: 0xA0,0x71
Write control
Command number: 7
Number of parameters: 3
Bytes sent by the GSV: 0
Write control
resets the configuration which was determined with
read control
before.
The type of operation, polarity, amplification and Notch-frequency (data frequency) are
set.
Affected register:
Channel, frequency, gain, polarity.
Value Range: 0x00.00.00..0xFE.76.FF
possible error codes: 0xA0,0x53,0x54,0x58,0x71
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf
22
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en
Summary of Contents for GSV-2AS