wrong parameter, functionality not implemented in this FW:
0x59
Not enough parameters or parameter timeout:
0x5A
Sending of parameters was impossible (unspecified):
0x90
(reserved)
Sending of parameters was impossible because tx-buffer was full:
0x91
Sending of parameters was impossible because (CAN-)Bus busy:
0x92
Execution was impossible because the receive buffer is full:
0x99
From this point: former command was eventually accomplished but:
Internal error (unspecified):
0x80
Intern arithmetical error:
0x81
Error at AD-converter settings
0x82
Error: measuring value unsuitable for desired action
0x83
EEPROM-Error (FW-Version 1.2.00 and above only)
0x84
* possibly; mainly additional parameters were loaded, that are not directly associated to this command. Also, no
additional parameters may have changed because the former settings and the loaded settings are similar.
** because of prohibited setting combinations
Set second threshold
Command number: 67
Number of parameters: 4
Bytes sent by the GSV: 0
Set second threshold
set the threshold of the amplifiers second threshold switch
(terminal No 13). The first two parameter bytes are the switch on threshold, the next two
bytes are the switch off threshold,both with the MSByte first. The switch on threshold
always needs to be bigger than the turn off threshold. You can change the threshold
functionality between hysteresis switch and window comparator
.
In the latter case,
the
values are to be interpreted as upper and lower switch thresholds.
Value Range: 0x00.01.00.00..0xFF.FF.FF.FE
Possible error codes 0xA0,0x56,0x71
Get second threshold
Command number: 68
Number of parameters: 0
Bytes sent by the GSV: 4
Get second threshold
determines the thresholds of the second threshold switch, as
described in
set second threshold
.
Possible error codes 0xA0,0x91
Get device Type
Command number: 69
Number of parameters: 0
Bytes sent by the GSV: 1
Get device Type reads the type number of the GSV, with the GSV-21 constantly 21d.
Possible error codes 0xA0,0x91
calc norm
Command number: 70
Number of parameters: 0
Bytes sent by the GSV: 0
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf Germany
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en
37