![ME-Messysteme GSV-2AS User Manual Download Page 31](http://html1.mh-extra.com/html/me-messysteme/gsv-2as/gsv-2as_user-manual_1757596031.webp)
Stop transmission
Command number: 35
Number of parameters: 0
Bytes sent by the GSV: 0
Stop transmission
stops the serial transmission of the measuring values. The serial
transmit buffer of the GSV will be emptied. This operation status will get lost after turning
off the amplifier.
Possible error codes 0xA0,0x71
Start transmission
Command number: 36
Number of parameters: 0
Bytes sent by the GSV: 0
Start transmission
starts the serial transmission of the measuring values, if they have
been stopped with „stop transmission“ before.
Possible error codes 0xA0
Clear buffer
Command number: 37
Number of parameters: 0
Bytes sent by the GSV: 0
Clear buffer
deletes the output buffer of the GSV.
Possible error codes 0xA0
Set mode
Command number: 38
Number of parameters: 1
Bytes sent by the GSV: 0
Set mode
configures the GSV for different modes of operation. As with most of the
operating parameters, this mode remains as it is even after turning the amplifier off.
Before you change particular bits of the mode-register you should to read it with
get mode
, change the bits as desired and write the byte back (read-modify-write). You
can only change bits 1..5. Description of the mode-variable:
MSB
Bit 6 Bit 5 Bit 4
Bit 3
Bit 2
Bit 1
LSB
blocking
x
AV-
Filter
Window c. mode Log-Mode
Max-Mode
Text-Mode
reserved
Text-mode
= 1:
measuring value transmission in ASCII-Format is active
Max-Modus
= 1:
maximum value transmission
is active
Log-Modus
= 1:
transmission of measuring values on request only is active
Window-Modus = 1:
threshold switch works as window comparator
AV-Filter
=1:
Adaptive averaging filter is on. Can only be enabled, if SpecialMode,FIR =0
Block:
=1:
Blocking-State: All Set- Write-commands will be rejected (see Switch
Blocking)
Read only
x:
do not change because of upwards compatibility
(read GSV mode before)
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf Germany
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en
31
Summary of Contents for GSV-2AS
Page 65: ......