GetAdaptFilterOptimize
Command number: 161
Number of parameters: 0
Bytes sent by the GSV: 2
With GetAdaptFilterOptimize the number of remaining measuring values during the self-
learning phase (as described above) can be read, until the self-learning phase is finished.
This is useful to determine when that phase will be finished. As the self-learning is finished
(or wasn't ever started), the answer to GetAdaptFilterOptimize is =0x0000; then the
measuring system can be used again normally.
Possible error codes: 0xA0,0x91
Read Ranges
Command number: 162
Number of parameters: 1
Bytes sent by the GSV: 4
With Read Ranges all configurable input sensitivities can be determined, this means even
those, who are currently not configured. The transfer parameter indicates which of
possible input sensitivities should be read. Here is High-Nibble (bits <7:4>) the channel
coded, which may be only 0 (bit 4 = 0) or 1 (bit 4=1). The lower nibble (bits <3:0>)
indicates whether the currently configured input sensitivity should be read (bits <3:0> =0)
or which of 3 existing input sensitivities levels. These may deviate from standard variant (1
/2 /3,5 mV/V) depending on the option of hardware.
The first byte of return is the socalled „Jumper-Group-ID“, which indicates (Bit4) =0 on the
channel nibble , for which jumper position applies the returned input sensitivity value.
The return value (bytes 2...4, High-endian) indicates the 100 time the input sensitivity in
mV (mV/V), which means it must be divided by 100, to get mV/V, or by 100000 to get
Volt.
Transfer parameter
Value
(Hex)
Meaning
Value at standard-version
(decoded, dec.)
0x00
Currently configured input sensitivity
1 / 2 / 3,5 mV/V / 10V
0x01
First input sensitivity for setting 1 of JP1 (s.S.Error:
reference not found)
1 mV/V
0x02
First input sensitivity for setting 2 of JP1
2 mV/V
0x03
Second input sensitivity for setting 2 of JP1
3,5 mV/V
0x10
Input sensitivity of channel 1 (if it is now activ)
10 V
0x11
Input sensitivity of channel 1
10 V