Get SensorText Assign
Command number: 168
Number of parameters: 0
Bytes sent by the GSV: 2
With Get SensorText Assign information about the assignment of settings to the saved
sensors and their names defined by user can be read.
The answer contains two parameters, each is 1 byte big. The first byte is the assignment
number. If it is =0, then there is no sensor(text) assigned to the settings. Otherwise it
describes the number (1 to 6) of user data record, in which the copy of the settings
belonging to sensor is taken (see Cmd 9/10: Get All / Save All).
The second byte contains flags about the presence of sensortext, s. table. If the bit is set,
there is a valid text, otherwise the bit is =0.
See also the next chapter “Using and managing of the user data record assigned text”.
Data record-Nr.
(frist byte of
answer)
Address range of
sensor text (s.
Cmd. 56/57), Hex
Address range of
sensor text (s. Cmd.
56/57), Dec
Bit-Nr. Second byte of answer. Bit =
1: text available
1
0x300...0x30D
768...781
Bit 0
2
0x310...0x31D
784...797
Bit 1
3
0x320...0x32D
800...813
Bit 2
4
0x330...0x33D
816...829
Bit 3
5
0x340...0x34D
832...845
Bit 4
6
0x350...0x35D
848...861
Bit 5
CANbus / CANOpen protocol of the GSV-2-CANOpen
Conformity: CiA 301 (communication) and CiA 404 (application)
Other documents: gsv2can.eds, ba-gsv2CanOpen.pdf
Connecting the CAN-Bus wires
The CAN-Bus wires are connected to the terminals D, E and F:
Terminal
Description
D
CAN-GND
E
CAN-L
F
CAN-H
The CAN-Interface is galvanically isolated.
Bus-termination
Just behind the 15-pole terminal block there's a 2-pole pin-connector for the bus-
termination. By closing these 2 pins with a jumper (1/10'' spacing) the bus-termination is
enabled.
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf Germany
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en
53