Set gage factor
Command number: 44
Number of parameters: 2
Bytes sent by the GSV: 0
Set gage factor
sets the gage factor. When using the GSV as a strain indicator, the
display scaling factor can be recalculated by the command
Calc Norm
, using the gage-
factor, the bridge type and the poissons ratio stored (the latter only if applicable). The
gage-factor register is transmitted as an integer value, as the real gage-Factor multiplied
by 100, with the MSByte first.
Value Range: 0x00.09..0x7F.BC (corresponding K-Factor= 0,09 to 327,00).
Possible error codes 0xA0,0x54,0x55,0x71
Get gage factor
Command number: 45
Number of parameters: 0
Bytes sent by the GSV: 2
Get gage factor
reads the gage-factor in the format described in
Set Gage factor
.
Possible error codes 0xA0,0x91
Set poisson
Command number: 46
Number of parameters: 1
Bytes sent by the GSV: 0
Set poisson
sets the poissons ratio µ. The poisson is important when calculating strain
gage bridge types with one or two measuring grids which are aligned crosswise. The
poisson register is transmitted as a whole byte, which is the poissons ratio multiplied by
500.
Value Range: 0x00..0xFA (correspondingly 0,000 to 0,500).
Possible error codes 0xA0,0x54,0x7
1
Get poisson
Command number: 47
Number of parameters: 0
Bytes sent by the GSV: 1
Get poisson
reads the poissons ratio as described in
Set Poisson
.
Possible error codes 0xA0,0x91
Set bridge type
Command number: 48
Number of parameters: 1
Bytes sent by the GSV: 0
Set bridge type
sets the bridge type of the strain gage sensor, with which the display
scaling factor can be recalculated by
Calc Norm
.
0: Full Bridge
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf Germany
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en
33