Figure 6-16 Lamp Assembly and Adjustment Plate
7.
Remove the three shoulder screws and their springs from the lamp. The springs are loosely
coiled around the screws and do not spring off the screw when removed. The lamp
adjustment plate can now be separated from the deuterium lamp.
Caution: Lamp gas is under slight pressure. Use care when disposing of the lamp to prevent
shattering the glass.
6.4.3 Installing the New Lamp
Unpack the lamp from its packing material. The new lamp may vary slightly from the lamp
illustrated in Figure 6-16.
Note: Do not touch the glass bulb on the new lamp. Dirt or fingerprints on the bulb affect detector
operation. If the lamp needs cleaning, gently clean the bulb with ethanol and lens tissue. Do
not use any abrasive tissue or excessive pressure.
To install the new lamp:
1.
Position the lamp housing on a flat surface so the large cut-out on the adjustment plate
faces away from you (Figure 6-17).
Lamp
aperture
Adjustment
plate
Shoulder
screws
Springed-shoulder
screws
Large cut-out
Lamp
aperture
Notch
Lamp connector