Caution: To prevent burn injury, do not touch the lamp assembly directly.
5.
Observe the displayed REF value while you turn the center adjustment screw
(see Figure 6-14) to maximize the REF value. Be sure the displayed value is stable before
continuing.
6.
If the screws on top of the lamp assembly (see Figure 6-14) are tightened, use an Allen-
head wrench to loosen them.
7.
Observe the displayed REF value while you rotate the lamp adjustment plate to maximize
the REF value.
8.
Turn off power to the CHDF 2000.
9.
Wait five seconds, then restart this procedure from step 1.
10.
Rotate the selector knob to set the wavelength (AU) to 656 (nm).
Caution: To prevent burn injury, do not touch the lamp assembly directly.
11.
Observe the displayed REF value while you turn the center adjustment screw
(see Figure 6-14) to maximize the REF value. Be sure the displayed value is stable before
continuing.
12.
If the screws on top of the lamp assembly (see Figure 6-14) are tightened, use an Allen-
head wrench to loosen them.
13.
Observe the displayed REF value while you rotate the lamp adjustment plate to maximize
the REF value.
14.
Repeat steps 11 through 13 to further maximize lamp energy.
15.
Turn off the instrument and check that all detector screws and connectors are secure.
16.
Replace the instrument covers and fasten the cover screws.