3-7
Matec Applied Sciences
56 Hudson St., Northborough, MA 01532 Ph: (508) 393-0155 Fax: (508) 393-5476
www.matec.com
e-mail: [email protected]
Figure 3.7 Instructions on how to prime pumps.
Note:
The order of the instructions shown in this screen is not correct. The correct sequence
of steps to follow are: 1, 2, 3, 6, 5, 4. The process will be explained here in words to avoid
confusion. The syringe attaches to a special valve in the lower right front of the instrument called
the priming valve. Attach the syringe to the priming valve, then open the valve. Draw several
mls of eluant into the syringe. Start the pump by selecting Press to START, remove the syringe
from the priming valve; wait until the pump stops, then close the valve. Please realize that this
screen only gives instructions, accessing this screen by itself will not prime the pumps.
3.2 Reviewing and Editing Operating Parameters
The CHDF 2000 is configured with default values for 14 operating parameters that control
specific system functions. Review all operating parameters and edit them if required to customize
instrument operation for your needs.
3.2.1 Changing Operating Parameters
To review and edit the system parameter values of the CHDF 2000 choose Configure system
from the Home Screen. The first of three Configure System screens appears (Figure 3.8).
Figure 3-8 First Configure System Screen