6-20
Matec Applied Sciences
56 Hudson St., Northborough, MA 01532 Ph: (508) 393-0155 Fax: (508) 393-5476
www.matec.com
e-mail: [email protected]
Figure 6-17 Aligning the Lamp (Top View)
2.
With the lamp aperture facing away from you, place the new lamp on top of the adjustment
plate and secure it with the shoulder screws. The lamp leads should line up with the notch
on the lamp adjustment plate.
3.
Without touching the bulb, turn the assembly so that the bulb is positioned downward.
4.
Place the center adjustment screw with the hex nut and star washer in the center opening of
the adjustment plate. In preparation for maximizing the lamp energy, turn the screw until
the lamp assembly moves about 3-4 mm away from the adjustment plate.
5.
Place the lamp assembly onto the lamp housing. Make sure the lamp leads slide freely into
the housing slot.
6.
Secure the lamp assembly to the housing and tighten the three Allen-head screws.
7.
Plug the lamp connector into the lamp power cable assembly (see Figure 6-14).
8.
Replace the protective shield to prevent direct access to the lamp when instrument power is
on.