Clearing System Blockages
1.
If you are analyzing latex or other polymeric based samples, a blockage may be removed by
pumping acetone or tetrahydrofuran (THF) through the system for 15 minutes. This will
hopefully dissolve the plug.
2.
Change (or clean) the in-line particle filter or its element (See section 6.6).
3.
You can also try backflushing the fractionation cartridge to remove a plug of material. The
idea is to pump eluant through the cartridge in a direction opposite to that which is normal.
To backflush the cartridge, open up the front door of the CHDF 2000. There is a panel on the
right hand side held in place with two screws. Remove these screws (they should come out
by hand), and you should see the following.
The idea is to swap the tubes at fittings 1, and 2, and separately 3 and 4 as follows: Disconnect
the stainless steel tube from Fitting 1. Connect the tube on Fitting 2 to Fitting 1; then connect the
tube from Fitting 1 to Fitting 2. Disconnect Fitting 3. Connect the tube on Fitting 4 to Fitting 3
and Fitting 3 to Fitting 4. Run the pump for 15 minutes. Reconnect all fittings as they are
normally, and try running the pump again. If the high pressure message does not re-appear, your
backflushing was successful.
4. Snip off end of the glass capillary tube in the bottom cartridge (see Section 6.6).
5. Replace the entire bottom fractionation cartridge (see Section 6.6).
Do not restart sample analysis until you correct the cause of high system pressure.
Press the selector knob to clear the High Pressure Limit Exceeded error message.
5.1.2 Low Pressure Limit Exceeded
The Low Pressure Limit error message (Figure 5-2) appears when the system pressure falls below
the low-pressure limit for more than 110% of the value defined for the Pump Priming Time
parameter. You configure the low-pressure limit using the Pressure Limit Set operating parameter
(refer to Chapter 3, Reviewing and Editing Operating parameters).