Eluant contaminated
Use fresh eluant.
Eluant not properly
degassed
Degas eluant.
Flow fluctuations
Correct pump
malfunctions. Replace
pump seals, check valves.
Section 6.1
Lamp Standby Delay too
short
Increase value of Lamp
Standby Delay parameter.
Section 3.2
Incorrect wavelength for
eluant
Change eluant or
wavelength.
UV absorbance
specifications of
eluant supplier
Baseline
drift, slow
Eluant contaminated
Use fresh eluant.
Decreased UV lamp
energy
Check lamp energy.
Section 5.3
Ambient temperature
fluctuations
Prepare the system to
proper equilibration.
UV detector flow cell
leaking
Check flow cell, tighten
connections.
Section 6.4