6-26
Matec Applied Sciences
56 Hudson St., Northborough, MA 01532 Ph: (508) 393-0155 Fax: (508) 393-5476
www.matec.com
e-mail: [email protected]
Figure 6.20 The Fractionation Cartridge. IMPORTANT NOTE: This is how your cartridge
should look upon uninstalling it from the CHDF unit. Do not remove the Inlet and Outlet tubes
until after this whole assembly has been removed from the CHDF unit. Otherwise, you will break
the capillary inside these two stainless steel tubes.
3. Disconnect the cartridge fitting from the instrument at locations 1, 2, 3, and 4 shown in Figure 6.20.
CAUTION: DO NOT LOOSEN FITTINGS NUMBER 5 and 6 YET. The fused-capillary stretches into the
two stainless steel tubes, 7 and 8. Detaching tubes 7 and 8 at this point will result in fractionation capillary
breakage.
Unscrew the three screws holding the cartridge in place.
4. Gently remove the cartridge from the instrument.
5. Remove the plastic caps from the tube ends of the new cartridge if they are present. Take the new
cartridge and place it gently into the instrument.
6. Replace the cartridge using the 3 screws provided.
7. Reconnect the cartridge tubing to the instrument. Do not over-tighten the fittings. Tighten ¼ turn past
finger-tight.
8. 1/16” Stainless Steel Outlet
Tube
2. Top Cartridge
Outlet
7. 1/16” Stainless Steel Inlet
Tube
3. Bottom cartridge
outlet
5. Capillary Outlet Fitting. Do not disconnect during cartridge removal/installation.
6. Capillary Inlet Fitting. Do
not disconnect during cartridge
removal/installation.
1. Capillary Inlet
4. Bottom Cartridge
Inlet.