www.masimo.com
85
Masimo
Chapter 5: NIBP Measurement
Root provides noninvasive blood pressure readings through an oscillometric method.
An oscillometric method of blood pressure measurement is a noninvasive method that
monitors the amplitude of cuff pressure changes during cuff deflation to determine arterial
blood pressure. The NIBP module is designed for use on adult, pediatric, and neonatal
patients.
Patient Measurement Mode
Below is a table that provides a method for selecting the appropriate NIBP patient category.
To change patient category see
Profiles
on page 53.
Weight
Patient Category Maximum Pressure
Greater than 75lbs (34kg)
Adult
280 mmHg
Between 15.4-75lbs (7-34kg) Pediatric
280 mmHg
Less than 15.4lbs (7kg)
Neonatal
140 mmHg
Cuff Selection and Placement
Root uses a quick connect hose with a blood pressure cuff to measure NIBP. Based on the Cuff
Type that is selected, use the following approved blood pressure cuff chart to choose sizes.
Accessory
Length
NIBP Patient Hose, Male Quick Connect 3 meters (9.8 feet)
NIBP Patient Hose, Female Bayonet
3 meters (9.8 feet)
Cuff Type
Size (color), Range - cm
Reusable Masimo Blood Pressure Cuff
Child (green), 12 - 19
Small Adult (turquoise), 17 - 25
Adult (navy blue), 23 - 33
Large Adult (burgundy), 31 - 40
Summary of Contents for Root
Page 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 10: ......
Page 20: ......
Page 80: ......
Page 86: ......
Page 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Page 104: ......
Page 106: ......
Page 108: ......
Page 112: ......
Page 116: ......
Page 136: ......
Page 146: ......
Page 150: ......
Page 151: ......
Page 152: ...www masimo com 38280 9274A 0316 ...