Root
Chapter 12: Specifications
www.masimo.com
119
Masimo
NIBP Pressurization Ranges
Weight
Patient
Category
Initial
Pressurization
Maximum
Pressure
Greater than 75 lbs (34 kg)
Adult
160 mmHg
280 mmHg
Between 15.4 - 75 lbs (7 - 34
kg)
Pediatric
140 mmHg
280 mmHg
Less than 15.4 lbs (7 kg)
Neonatal
90 mmHg
140 mmHg
Electrical
Root
AC Power requirements 100-240 VAC
~
, 47-63 Hz, 180 VA (Max)
Fuses (2)
2 Amp, Metric, (5x20mm), 250V, 1500A Breaking Capacity
Battery
Type
10.8V Lithium Ion (Nominal)
Capacity
4 hours*
Maximum Charging Time 4 hours
*This represents approximate run time at the lowest brightness, using a fully charged battery.