Chapter 3: Operation
www.masimo.com
51
Masimo
IntelliBridge Connectivity
IntelliBridge connectivity allows Root to transmit parameters and waveforms to Philips
multi-parameter patient monitors that support Philips IntelliBridge device interfacing
modules. This option allows parameters and waveforms on Root to be displayed on a Philips
monitor and, if applicable, transmitted to the electronic medical record system.
Masimo parameter and waveforms from SET®, rainbow SET® and SedLine® channels are
supported.
Note
: Root supports the transmission of data only. Validations of the retrieval and display of
data transmitted is the responsibility of the IntelliBridge manufacturer.
Parameters Supported
IntelliBridge connectivity allows for up to six (6) parameters and two (2) waveforms or eight
(8) parameters and no waveform to be displayed on Philips monitors.
Channel
Supported Parameters
Waveforms
SET®
SpO2, PR, PI, PVI
Pleth
rainbow® RRa®, SpHb®, SpCO®,SpOC™,SpMet® RR
SedLine PSI™, SR, EMG, ARTF, SEFR, SEFL
(EEG) L1, L2, R1, R2, L1L2, R1R2