www.masimo.com
117
Masimo
Chapter 12: Specifications
This chapter contains specifications of Root.
For information on the specifications of Radical-7, Radius-7, MOC-9 modules, and standalone
devices, see Directions for Use or Operator's Manuals for these devices.
Measurement Accuracy
NIBP
Pressure Transducer
Between 0 mmHg and 300 mmHg ±3mmHg
Temperature*
Temperature
Between 80°F and 110°F
(26.7°C and 43.3°C)
All patient populations
±0.2
°F (0.11°C)
Temperature
Measurement Site/Mode
Number of
Subjects
Clinical
Bias (°C)
Limits of
Agreement
Clinical
Repeatability
Oral
106
0.01
0.63
0.14
Rectal
105
-0.12
0.59
0.29
Pediatric Axillary
117
-0.03
0.56
0.14
Adult Axillary
105
0.13
0.43
0.14
*Test Report #:2010036 Test Report of Clinical Investigation to demonstrate that the SureTemp® Plus thermometer meets the essential
performance of clinical thermometers for body temperature measurement as described in ISO 80601-2-56:2009