Chapter 3: Operation
www.masimo.com
58
Masimo
Alarms for Temperature
From the
Temperature
screen, touch
Alarms
, and then change any of the following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
Configurable
Options
High Limit
(°F)
The High Limit is upper
threshold that triggers an
alarm.
High
Off
80.20-109.9 in
steps of 0.1, or Off
When set to Off,
alarm is disabled
Low Limit
(°F)
Low Limit is the lower
threshold that triggers an
alarm.
High
Off
80.1-109.8 in steps
of 0.1, or Off
When set to Off,
alarm is disabled
High Limit
(°C)
The High Limit is upper
threshold that triggers an
alarm.
High
Off
26.9-43.2 in steps of
0.1, or Off
When set to Off,
alarm is disabled
Low Limit
(°C)
Low Limit is the lower
threshold that triggers an
alarm.
High
Off
26.8-43.1 in steps of
0.1, or Off
When set to Off,
alarm is disabled
Silence
Duration
Temporarily suspend
audible alarms for a period
of time.
None
2 min
30 sec, 1 min, 2 min,
5 min
Summary of Contents for Root
Page 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 10: ......
Page 20: ......
Page 80: ......
Page 86: ......
Page 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Page 104: ......
Page 106: ......
Page 108: ......
Page 112: ......
Page 116: ......
Page 136: ......
Page 146: ......
Page 150: ......
Page 151: ......
Page 152: ...www masimo com 38280 9274A 0316 ...