Chapter 11: Troubleshooting
www.masimo.com
116
Masimo
Symptom
Possible Cause
Correction
Iris screen does not
display connection
status for standalone
devices.
Connection error.
If the problem persists, refer to instructions for
use or operator's manual for the connected
standalone devices or Iris section of the
instructions for use or operator’s manual for
the appropriate version of Patient SafetyNet.
Root has a continuous
speaker tone.
Internal failure.
To silence an alarm, press the Power Button for
eight (8) seconds. If alarm continues to sound,
Power Off Root. Root requires service. See
Return Procedure
on page 140.
Power Button does not
respond when
pressed.
Power Button may
need to be pressed
for a longer time.
To Power On when turned off or in Sleep Mode,
press Power Button for two (2) seconds.
To Power Off when turned on or in Sleep Mode,
press the Power Button for eight (8) seconds.
Internal failure.
Root requires service. See
Return Procedure
on
page 140.
Home Button does not
work when pressed.
Internal failure.
Root requires service. See
Return Procedure
on
page 140.
Battery does not
charge.
AC power cable may
be disconnected.
Unplug and replug AC power cable.
Root Charging
Indicator illuminates
red.
Internal failure.
Root requires service. See
Return Procedure
on
page 140.
Nurse Call does not
communicate.
Connection error.
Unplug and replug Nurse Call connector.
See
Nurse Call Setting Connections
on page
Summary of Contents for Root
Page 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 10: ......
Page 20: ......
Page 80: ......
Page 86: ......
Page 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Page 104: ......
Page 106: ......
Page 108: ......
Page 112: ......
Page 116: ......
Page 136: ......
Page 146: ......
Page 150: ......
Page 151: ......
Page 152: ...www masimo com 38280 9274A 0316 ...