Chapter 4: Temperature Measurement
www.masimo.com
80
Masimo
2.
To change the measurement site, touch the
Main Menu
icon
, then select
Temperature Settings
.
Select
Additional Settings
, then select the desired measurement site
through
Probe Mode
. The measurement site can also be changed using the action
menu.
3.
Remove the temperature probe from the probe well holder to initiate spot check.
Note:
Ready status will display in the status bar and sound will play.
4.
Apply a single-use disposable probe cover to the temperature probe before
measurement.
WARNING:
The use of any other probe cover may produce temperature
measurement errors or may result in inaccurate readings.
5.
Place the temperature probe on the site to begin spot check.
WARNING:
Before use, verify the color of the probe cover eject button on the
temperature probe to confirm the proper application site: Red (rectal), Blue
(oral/axillary).
Note:
Measuring status displayed in the status bar.