Root
Chapter 3: Operation
www.masimo.com
50
Masimo
Device Output
A Nurse Call can be triggered based on alarm, low Signal IQ events or both. In addition, Nurse
Call Polarity can be inverted to accommodate local Nurse Call station requirements.
Option
Description
Factory
Default
Setting
Configurable Settings
Nurse Call
Trigger
Controls the source of monitoring
which sets off the trigger.
Alarms
Alarms, Signal IQ
Alarms, and Signal IQ
Nurse Call
Polarity
Controls the mechanism of action for
triggering to occur. Should be
changed to accommodate
institutional Nurse Call settings.
Normal
Normal or Inverted
USB Port 1
Controls the output type for USB Port
1.
None
None, IAP,
IntelliBridge, and ASCII
1
USB Port 2
Controls the output type for USB Port
2.
IAP
None, IAP,
IntelliBridge, and ASCII
1
IntelliBridge
Output
Controls the data type output for
IntelliBridge.
Radical
Radical, Radical Module
A, SedLine, and SedLine
Numerics Only
Note: The Nurse Call feature is disabled when Audio Pause is enabled and Nurse Call Trigger
is set to
Alarms
. For more information about Audio Pause, see
Audio Pause
on page 71.
Summary of Contents for Root
Page 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 10: ......
Page 20: ......
Page 80: ......
Page 86: ......
Page 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Page 104: ......
Page 106: ......
Page 108: ......
Page 112: ......
Page 116: ......
Page 136: ......
Page 146: ......
Page 150: ......
Page 151: ......
Page 152: ...www masimo com 38280 9274A 0316 ...