Chapter 11: Troubleshooting
Troubleshooting Radical-7, Radius-7, and MOC-9 Modules
For information on troubleshooting values that are provided from Radical-7, Radius-7, and
MOC-9 modules, refer to their respective Instructions for Use or Operator's Manuals.
Troubleshooting Root
Symptom
Possible Cause
Correction
Root does not turn on.
Power Button not
pressed long
enough.
Press Power Button for two (2) seconds.
The battery may be
depleted.
Connect Root to AC power to charge battery.
One of the fuses is
not operating
properly.
Replace the fuse. See
Replacing the Fuses
on
page 135.
Root turns on, but
Main Screen is dim or
blank.
The brightness
setting is not
correct.
Adjust the brightness setting. See
Brightness
on page 48.
If the condition persists, Root requires service.
Contact Masimo Technical Support. See
Return Procedure
on page 140.
Touch functionality is
not responsive.
Internal failure.
Root requires service. Contact Masimo
Technical Support. See
Return Procedure
on
page 140.
Not displaying data
from Radical-7,
Radius-7, or MOC-9
modules.
Connection error.
Ensure that the connections are securely in
place and properly plugged in, or that the
cable is not defective. For Radius-7, ensure
that the device is paired with Root via
Bluetooth. Refer to Operator's Manual of
Radius-7 for more information.
Iris screen does not
display connection
status for standalone
devices.
Connection error.
Unplug and replug the Iris Adapter.
Summary of Contents for Root
Page 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Page 2: ......
Page 4: ......
Page 8: ......
Page 10: ......
Page 20: ......
Page 80: ......
Page 86: ......
Page 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Page 104: ......
Page 106: ......
Page 108: ......
Page 112: ......
Page 116: ......
Page 136: ......
Page 146: ......
Page 150: ......
Page 151: ......
Page 152: ...www masimo com 38280 9274A 0316 ...