69
Note 2:
Protected against splashing water.
Water splashed against the enclosure from any direction
shall have no harmful effects
Transient
operating
conditions:
temperature
-4°F to 122° F; -20°C to 50° C
Transient
operating
conditions:
humidity
15% to 95% relative humidity,
the time required for the
ME EQUIPMENT to
warm from the minimum
storage temperature
between uses (4.2.1) until
the ME EQUIPMENT is
ready for its INTENDED
USE when the ambient
temperature is 20 °C
<6 hour
the time required for the
ME EQUIPMENT to cool
from the maximum storage
temperature between uses
(4.2.1) until the ME
EQUIPMENT is ready for
its INTENDED USE when
the ambient temperature is
20 °C
<1min
10.3
Defibrillator
Category
Specification