11
2.2.5
Notices
If the battery pack is removed for any reason, the user must
label the AED7000 Plus as "
Out of service due to battery
operation.
"
Do not use mobile phones near the AED7000 Plus as the
radiation field produced by mobile phones will interfere with
the functioning of the AED7000 Plus.
A warning that the AED7000 Plus should not be used
adjacent to or stacked with other equipment and that if
adjacent to or stacked use is necessary, the AED7000 Plus
should be observed to verify normal operation in the
configuration in which it will be used.
2.2.6
Attention
Electric and magnetic fields can interfere with the
equipment’s regular work. Therefore, please ensure all
external devices used near the AED7000 Plus comply with
EMC demands.
2.3
Patient Safety
2.3.1
General
Warning
Accessory adapter required for use on
pediatric patients.
he AED 7000 is not to be used on patients less
than 8 years old or under 25kg
(55 lbs), unless it is equipped with the