45
socket of the AED7000 Plus, push the connector firmly
into the socket.
6.3.2
Troubleshooting of analysis
Problem description:
The analysis is interrupted during the analysis period, and the
voice prompt of “Check the electrode” is broadcast.
Solution:
a)
Shut down the AED7000 Plus and check the pads
placed on the patient, if they are not properly connected,
press the pads firmly on the chest of the patient, then
start the AED7000 Plus again.
b)
If the patient has moved during the analyzing period,
shut down the AED7000 Plus and immobilize the
patient if possible.
6.3.3
Troubleshooting of charging
Problem description:
The charging is stopped during the charging period, and the voice
prompt of “Charging cancelled: check the electrode” is broadcast.
Charging cancelled: check the electrodes.
Solution:
a)
Shut down the AED7000 Plus and check pads
connector. If the connector inadequately connects to the
socket of the AED7000 Plus, push the connector firmly