12
accessory AED 7000 Pediatric Energy
Reducer.
Caution
Patient Physical Harm
Place the AED7000 Plus in a position where it
cannot harm the patient should it fall. Do not
use adjacent to or stacked with other
equipment. Keep all cables and connectors
away from the patient's neck.
2.3.2
Shock Hazard
Warning
Defibrillation current can cause injury
Do not touch the patient during defibrillation.
Do not touch equipment connected to or metal
objects in contact with the patient during
defibrillation.
Disconnect other electrical equipment which
has no DEFIBRILLATION-PROOF applied
parts from the patient before defibrillating.
Avoid contact between parts of the patient’s
body such as exposed skin of head or limbs,
conductive fluids such as gel, blood or saline
and metal objects such as a bed frame or a
stretcher which may provide unwanted
pathways for the defibrillating current.