37
seconds.
Time out, analysis restart.
5.8
Cardiopulmonary resuscitation procedure
(CPR)
The AED7000 Plus will enter cardiopulmonary resuscitation
procedure, when the condition below occurs:
Note
During the CPR procedure, the four indicators: Check
Pad, Caution, Analyzing and Shock will not flicker.
No shockable heart rhythm is detected
During the heart rhythm analysis period, if the heart rhythm is not
considered to be a shockable rhythm by the AED7000 Plus, a
cardiopulmonary resuscitation procedure (CPR) will start.
No shock advised
During the period of charging, if the AED7000 Plus detects that
the shockable rhythm has changed to a normal one, the AED7000
Plus will stop current procedure and switch to CPR procedure.
1.
The operator can touch the body of the patient and check their
conditions.
Note
During the cardiopulmonary resuscitation period, the
AED7000 Plus will ignore the heart rhythm analyzed
result and the improper electrode connection.