14
Warning
Properly place defibrillation pads
Do not place electrodes near the generator of an
internal pacemaker. Always apply electrodes
to flat areas of skin. Avoid application over
folds of skin such as those underneath the
breast or on obese patients. Excessive hair,
poor adhesion, or air under electrode may
produce burns.
2.3.5
ECG Misinterpretation
Warning
Properly place defibrillation pads
Improperly placed pads may produce incorrect
analysis and an inappropriate shock or no shock
advisory decision.
Warning
Do not move patient
Handling or transporting the patient during
ECG analysis can cause incorrect or delayed
diagnosis. Follow all instructions in the Users
Manual.
Warning
Cardiac pacemakers may affect rhythm
analysis
Patient pacemakers may reduce the sensitivity
of the AED7000 Plus analysis and errors in
detecting shockable rhythms.
Warning
Radio frequency (RF) interference
Do not operate the AED7000 Plus in
conjunction with electro-cautery or diathermy
equipment. Any equipment that emits strong