46
into the socket.
b)
Shut down the AED7000 Plus and check the pads
contacting with patient’s bare chest. If the pads are
poorly contacted, wipe off any moisture and press the
pads firmly on the chest of the patient.
6.3.4
Troubleshooting of defibrillation
Problem description:
The AED7000 Plus automatically has an internal disarm function
following completion of the analysis process. The unit
broadcasts a voice prompt of “Time out, analysis restart”.
Time out, analysis restart.
Solution:
If the operator doesn’t press the shock button in 30
seconds after the analysis is finished, the AED7000 Plus
will automatically disarm.
Press the shock button in 30 seconds after the voice
prompts of “press flashing shock button”.
Press flashing shock button.
6.3.5
Troubleshooting of battery
Problem description:
During the operation of the AED7000 Plus, the ON/OFF indicator
is flickering.