40
6
Maintaining the Automatic
External Defibrillator
The maintaining, cleaning and Troubleshooting of the AED7000
Plus will be described in this section:
Maintaining
Cleaning and disinfecting
Troubleshooting
6.1
Maintenance
Maintenance is required for the optimal operation of the AED7000 Plus.
Maintenance should be performed periodically; with weekly or monthly
inspections recommended.
The frequency of inspection:
It is recommended that the AED7000 Plus is inspected once a
week or twice a month, but if use of the AED7000 Plus is more
frequent, additional inspections should be performed.
The table below describes a general rule to perform the
inspection:
Frequency of use (n / month)
inspection
1 / month
Once a week
2 ~ 3 / month
Twice a week
Infrequent using(1 ~ 2 a year)
Once a month
Caution
If the AED7000 Plus should be used more than 4 times
a month, we recommend that the AED7000 Plus
should be inspected by authorized service personnel at
least twice a year.