Patient cable connector
3.4.2
Controls
The AED7000 Plus is designed for ease of operation. After putting
the defibrillator pads on the patient and connecting them to the
AED7000 Plus, the operator performs this simple three-step
process:
1. Turn the power ON.
2. Follow LED prompts on the panel and voice prompts from the
speaker.
3. If prompted, deliver shock by pressing the flashing red Shock
button.
Power On/Off
Green On/Off button to toggle system power
on/off.
Shock
Red Shock button to discharge defibrillator,
red LED flashes when defibrillator is fully
charged.
3.4.3
LCD Prompts
Operating information and user instructions display in these
light-emitting diodes.
Display
information
Description
Check Pad LED
Flashes
Defibrillator pads are not properly
attached to the patient or are not properly
connected to the AED7000 Plus.
Charging ,do not
touch the patient
Do not touch or move the patient.
Analyzing do not
touch the patient
ECG leads are properly connected and
the system is assessing the patient's heart