77
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE. U
T
is the AC mains voltage prior to application of the test level.
10.6.3
Electromagnetic immunity (IEC 60601-1-2 Table
203)
The AED7000 Plus is intended for use in the electromagnetic
environment specified below. The customer or the user of the
AED7000 Plus should ensure that it is used in such an
environment.
Immunit
y test
IEC
60601
test level
Complia
nce level
Electromagnetic environment –
guidance
Radiated
RF IEC
61000-4-
3
10V/m 80
MHz to
2.5 GHz
20V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the
AED7000 Plus, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance