Troubleshooting
As with any equipment, problems may occur during the operation
of the AED7000 Plus. Indicators and voice prompts will be given
to the operator to indicate any identified issues with the unit.
General trouble shooting tips are noted below.
Note
The operator could try to salve the problem which occurring prior
to or during the operation according to these trouble shootings
beblow
6.3.1
The troubleshooting of attaching electrode pads
Problem description:
The voice prompt of “Check the electrode” continues to broadcast
when the electrode pads have been placed on the patient and the
pads connector has been connect to the socket of the AED7000
Plus.
Solution:
a)
If the electrode pads are dry, damaged or out-of-date,
replace the current pads with new ones and
recommence.
b)
Ensure that the electrode pads have been properly
adhered to the patient. If not attached correctly, wipe off
any water or moisture from the chest and shave the hair
from chest area, then press the pads firmly on the chest
of the patient.
c)
Ensure the two pads do not touch each other.
d)
If the pads connector is inadequately connected to the
Check the electrodes.