Charging is canceled
The system will broadcast this audio if some unusual condition occurs in the
charging module, and release the energy in charging capacity then reenter the
ECG analyzing procedure.
8.
Charging failed
The system will broadcast this audio if the charging module can not load the
expected energy into the charging capacity, in such a case, the system releases the
energy in charging capacity then reenter the ECG analyzing procedure.
9.
Time out
When the system enters the “waiting for discharging” procedure, the system will
be blocked and wait for the operator to press the shock button to externally
release the energy to the patient, but this audio will be broadcast if the operator
do not press the shock button before the waiting time expires, the system will
internally release the energy in the charging capacity then reenter the ECG
analyzing procedure.
10.
Press the shock button
This audio will be broadcast when the end of successful charging, On hearing this,
the operator must press the shock button to carry out the cure to the patient.
11.
Waiting is canceled
If some ususual condition occurs in the“waiting for discharing”procedure, the