Section 4: Power-up Defaults Menu and Diagnostic Mode
4-5
•
System Software Version
: Displays the revision level of the system
software. The revision level is also momentarily shown on the LCD as part
of the Copyright screen. This value may not be changed by the user.
•
SpO
2
Software Version
: Displays the revision level of the software of the
MP-205 SpO
2
module. This value may not be changed by the user.
When in the System Information screen, the knob may be rotated to select any of
the “changeable” items. If one of those items is selected, a press of the knob will
cause a pop-up menu to appear. The first item in the pop-up will read “Make no
change”; the second item in the pop-up will read “Reset to zero”. Exiting the
screen is accomplished in the normal manner, by selecting “Return”.
4.3.3 System A/D Values
This screen displays the current value of each analog-to-digital (A/D) channel, in
volts. Some of the channels are for AC-coupled signals (such as ECG input), so
the numbers on the screen will be constantly changing when an input signal is
present. These AC-coupled values are shown to give an indication as to whether
basic functionality of the channel is present, but no significance can be derived
from the values of the numbers displayed. However, others of the A/D channels
read DC voltages, (for example, power supply voltages and battery voltage)
those voltage values directly provide useful diagnostic information.
The Primary and Secondary Status messages from the SpO
2
module will be
displayed and updated at the rate of about once per second. Presence of the
correct SpO
2
message indicates that, at a basic level, communication between
the SpO
2
module and the main monitor processor is working correctly. None of
the displayed values may be changed or reset in this screen.
When in the System A/D screen, the “Return” item is always highlighted; a press
of the knob will return the user to the Diagnostic Menu. Rotating the knob while
in the System A/D screen will have no effect . The A/D channel designators are
shown in Table 4-2.
Table 4-2: A/D Channel Designators
A/D CHANNEL DESIGNATOR
A/D CHANNEL DESIGNATOR
1.
ECG
12.
(BATTERY VOLTAGE) X 0.5
2.
RWAVE
13.
not used
3.
PACEMAKER
14.
+3.3VDC POWER SUPPLY
4.
RESPIRATION
15.
(+12VDC POWER SUPPLY) X
0.33
5.
PRESSURE XDUCER 1
16.
(NIBP VOLTAGE REF) X 0.8
6.
PRESSURE XDUCER 2
17.
GROUND REFERENCE
7.
NIBP OSCILLATORY
18.
(+5 VDC POWER SUPPLY) X 0.8
8.
ECG LEADS OFF
19.
ADC MID-SCALE VALUE
9.
TEMERATURE
20.
ADC FULL-SCALE VALUE
10.
ISOLATED VOLTAGE REF
21.
ADC ZERO-SCALL VALUE
11.
ISOLATED VOLTAGE ZERO
SpO
2
S1 S018
SpO
2
S2 S010
Summary of Contents for NELLCOR NPB-4000
Page 66: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 68: ...Section 7 Spare Parts 7 2 Figure 7 1 NPB 4000 C Top Assembly Drawing ...
Page 70: ...Section 7 Spare Parts 7 4 Figure 7 2 NPB 4000 C Front Case Assembly Diagram Sheet 1 of 2 ...
Page 72: ...Section 7 Spare Parts 7 6 Figure 7 3 NPB 4000 C Front Case Assembly Diagram Sheet 2 of 2 ...
Page 74: ...Section 7 Spare Parts 7 8 Figure 7 4 NPB 4000 C Rear Case Assembly Diagram Sheet 1 of 2 ...
Page 76: ...Section 7 Spare Parts 7 10 Figure 7 5 NPB 4000 C Rear Case Assembly Diagram Sheet 2 of 2 ...
Page 78: ...Section 7 Spare Parts 7 12 Figure 7 6 NPB 4000 C Power Supply Heat Sink Assembly Diagram ...
Page 80: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 96: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 114: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 140: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 180: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 192: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 208: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 210: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 211: ...Section 17 Drawings 17 3 Figure 17 1 MP 205 PCB Schematic Sheet 1 of 2 ...
Page 212: ...Section 17 Drawings 17 5 Figure 17 2 MP 205 PCB Schematic Sheet 2 of 2 ...