Section 3: Performance Verification
3-10
3.3.4.4 Verification of Pneumatic System
Tests in paragraphs 3.3.4.4.1 through 3.3.4.4.5 verify the functionality of the
NPB-4000/C pneumatic system. These tests are designed to use the Bio-Tek
“BP Pump” noninvasive blood pressure simulator. The internal test volume of
the Bio-Tek simulator is 250 cm
3
, which is used to calculated the
inflation/deflation rate periods. The Bio-Tek simulator or any equivalent NIBP
simulator is required to perform these tests.
The NPB-4000/C must be placed in Diagnostic Mode, with the NIBP Test screen
active for each of the NIBP tests. For a detailed explanation of the Diagnostic
Mode, refer to Section 4,
Power-up Defaults Menu and Diagnostic Mode
.
Each of the tests described in paragraphs 3.3.4.4.1 through 3.3.4.4.5 must be
performed to verify pneumatic system functionality. These tests can be
performed individually (in any order) or sequentially. Prior to performing any of
these tests, perform the following setup procedure. If these tests are performed
in sequence, this procedure needs to be performed once prior to the first test.
1. Turn on Bio-Tek simulator and press MODE button to place simulator in
test mode. Simulator screen will indicate “Internal Cuff” and “Pressure
Gauge”.
2. Connect simulator hose to NIBP connector on NPB-4000/C.
3. Follow procedure described in Section 4 to place NPB-4000/C in Diagnostic
Mode with NIBP Test screen active.
3.3.4.4.1 Pressure Transducer Accuracy
The pressure transducer accuracy test verifies the pressure accuracy of the
NBP-4000/C pressure transducer.
1. Ensure Bio-Tek simulator is in test mode. Simulator should display
“Pressure Gauge”.
2. Ensure simulator is set up for internal cuff.
3. Ensure NIBP Test screen is active on NPB-4000/C.
4. Press Alarm Silence switch on NPB-4000/C to ensure both valves are
closed.
5. Perform offset adjustment so that simulator and NBP-4000/C both display a
pressure of 0 mmHg.
a.
Press Contrast Adjust switch on NPB-4000/C,
b. Press Zero button on the simulator.
Summary of Contents for NELLCOR NPB-4000
Page 66: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 68: ...Section 7 Spare Parts 7 2 Figure 7 1 NPB 4000 C Top Assembly Drawing ...
Page 70: ...Section 7 Spare Parts 7 4 Figure 7 2 NPB 4000 C Front Case Assembly Diagram Sheet 1 of 2 ...
Page 72: ...Section 7 Spare Parts 7 6 Figure 7 3 NPB 4000 C Front Case Assembly Diagram Sheet 2 of 2 ...
Page 74: ...Section 7 Spare Parts 7 8 Figure 7 4 NPB 4000 C Rear Case Assembly Diagram Sheet 1 of 2 ...
Page 76: ...Section 7 Spare Parts 7 10 Figure 7 5 NPB 4000 C Rear Case Assembly Diagram Sheet 2 of 2 ...
Page 78: ...Section 7 Spare Parts 7 12 Figure 7 6 NPB 4000 C Power Supply Heat Sink Assembly Diagram ...
Page 80: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 96: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 114: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 140: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 180: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 192: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 208: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 210: ... THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 211: ...Section 17 Drawings 17 3 Figure 17 1 MP 205 PCB Schematic Sheet 1 of 2 ...
Page 212: ...Section 17 Drawings 17 5 Figure 17 2 MP 205 PCB Schematic Sheet 2 of 2 ...