80
DL30-G USERS MANUAL EM-8571-G Rev.2
Assigning binary accumulation to PI
The cumulative total is calculated by considering AI input values as a 16 bit unsigned integer and calculated
value is obtained as a 32 bit unsigned integer.
(1) Click [Pulse input (PI)] button in the [Input/Output] window to display the [Pulse input (PI)].
Double-click a row of the PI channel to be set in this window to display the [PI setting] window.
(2) Set the [CH setting] as [Binary accumulation], and enter the [AI CH No.].
1. Select
2. Enter
Parameter
Description
AI CH No.
Select AI CH to use for binary accumulation.
NOTES
Make sure that the AI range is set between 0 – 65535 and the pulse count is reset to 0 when
a count overflow occurs. Note that the DL30-G resets the pulse count to 0 also when the cur-
rent value is smaller than the previous value.
Input/Output
Pulse input (PI)
1. Click
PI setting
2. Double-click