| Operator’s Manual
47
S E C T I O N 9 • O p e r a t i n g I n s t r u c t i o n s
WARNING!
Blood and blood products could contain pathogenic organisms.
Failure to follow Institutional policy and procedures for
biomedical-hazardous materials could lead to exposure to
harmful pathogens which could result in user death or serious
injury.
Step 3—Make Patient Connection
Make patient connection and begin infusion.
Step 4—Replace the Gas Vent/Filter Assembly
Replace the Gas Vent /Filter Assembly every 3 hours, when the
filter becomes clogged, or if air is venting slowly. Refer to Section
9.3,
Replace the Gas Vent/Filter Assembly.
Step 5—Change the Fluid Bag
a
Move the lever on the Pressure Chamber over to the
minus (–) unpressurized position. This will release the pressure in
the Pressure Chamber and deflate the bladder.
Close pinch clamp under empty bag.
b
Open door and remove the fluid bag from the Pressure Chamber.
c
Remove the spike from the used fluid bag.
d
Remove any air from the new fluid bag and spike the fluid bag.
e
Hang the new fluid bag in the Pressure Chamber. Close and latch
the door.
f
Move the lever on the Pressure Chamber over to the
plus (+) pressurized position to pressurize the chamber.
g
Open pinch clamp.
Summary of Contents for H-1200 115V
Page 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Page 2: ...ii Operator s Manual...
Page 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Page 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Page 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Page 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Page 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...